Sample
Name
CO2 snow with 0.10% of a PAH mix
Date
2017-10-17
Mass
31.0327 $\pm$ 1.5001 $g$
Sample changes
$CO_2$ sublimation: sample thickness decreased from 18 to 15 mm during spectrum
Substrate material
aluminium
Substrate comments
cylindrical 48mm diameter, 20mm deep
Comments
sublimating sample with an initial thickness of 20 mm decreasing with time
already 2 mm CO2 sublimated at start of spectrum - possible increase of PAH concentration
Temperature
173.0 $\pm$ 2.0 $K$
Time at T
0.5 $\pm$ 0.5 $h$
Comments
sample in cell about 30min before first spectrum, not taken into account in 'time'. Full spectrum takes about 1h
Type
molecular gas
Fluids
Fluid temperature
173.0 $\pm$ 2.0 $K$
Fluid pressure
6.0 $\pm$ 0.5 $mbar$
Title
preparation of a mixture of PAH and CO2 snow

Precursors

Matters

Produced sample

Sample
CO2 snow with 0.10% of a PAH mix (this sample)
Processing steps
Step Chronology Date Type Process Changes
#1 before layer formation mixing mixing a weighted amount of PAHs mix with CO2 ice in a liquid N2 cooled stainless steel bowl the final mixture of CO2 ice and PAH is weighted to determine the PAH mass fraction
#2 before layer formation layer formation filling the liq N2 cooled cell with the PAHs/CO2 ice mixture, plus some manual densification and surface flattening flat homogeneous PAHs/CO2 ice mixture sample
Comments
some very thin and very fine grained layer of water ice may condense on the sample during closure of the cryogenic cell