- Title
- Pt L3 edge XAS fluorescence and XAS transmission of Aqueous platinum chloride soutions at 600-700 nars and 30°C
- DOI
- 10.26302/SSHADE/EXPERIMENT_GP_20180208_001
- Data reference
- Pokrovski, Gleb; Hazemann, Jean-Louis (2018): Pt L3 edge XAS fluorescence and XAS transmission of Aqueous platinum chloride soutions at 600-700 nars and 30°C. SSHADE/FAME (OSUG Data Center). Dataset/Spectral Data. https://doi.org/10.26302/SSHADE/EXPERIMENT_GP_20180208_001
- Database(s)
- Experimentalists
- Type(s)
- laboratory measurement
- Description
- Platinum (IV) and platinum (II) chloride aqueous solutions references at concentration around 0.1m Pt, in transmission and in fluorescence.
- Number of spectra
- 6
- Comments
- These data were acquired on the BM30 beamline in the frame of the 30-02-1131 proposal.
- Variable type(s)
-
- sample composition
- observation mode
- Instrument
- X-ray Absorption Spectroscopy measured in fluorescence mode
- Sample holder
- Institut Néel autoclave, see Testemale et al., Rev. Sci. Instrum., 2005.
- Standard medium
- unknown
- Observation mode
- spectrum
- Spectral range type(s)
- hard X
- Valid spectral range(s)
-
Min - Max ($eV$) Sampling ($eV$) Resolution ($eV$) Position accuracy ($eV$) Absorption edge #1 11446.9 - 12303.8 0.593 $Pt$ (L3: 11.5637 $keV$) - Comments
- The resolution in the above table is the resolution given by the monochromator. The energy was calibrated by measuring the absorption edge of a reference: Pt metallic foil (first maximum of the derivative set at 11564 eV).
Definition: emergence angle is positive with origin at nadir. Azimuth origin (increasing clockwise) is arbitrary.
- Observation geometry
- directional
- Observation mode
- fixed angles
- Comments
- Emitted x-rays are collected in a solid angle directed at 90 degrees from the incident beam.
- Observation mode
- averaged
- Instrument
- X-ray Absorption Spectroscopy measured in transmission mode
- Sample holder
- Institut Néel autoclave, see Testemale et al., Rev. Sci. Instrum., 2005.
- Standard medium
- unknown
- Observation mode
- spectrum
- Spectral range type(s)
- hard X
- Valid spectral range(s)
-
Min - Max ($eV$) Sampling ($eV$) Resolution ($eV$) Position accuracy ($eV$) Absorption edge #1 11446.9 - 12303.8 0.593 $Pt$ (L3: 11.5637 $keV$) - Comments
- The resolution in the above table is the resolution given by the monochromator. The energy was calibrated by measuring the absorption edge of a reference: Pt metallic foil (first maximum of the derivative set at 11564 eV).
Definition: incidence and emergence angles are positive with origin at nadir. Emergence angle in transmission is larger than 90° (=180°-i).
- Observation geometry
- direct
- Observation mode
- fixed angles
- Observation mode
- averaged
- Date begin - end
- 2018-02-08 - 2018-02-18
Versions
- Release date
- 2022-12-13 11:19:20 UTC+0000
- Version (Date)