Experiment
Title
Vis-NIR reflectance spectra (i=0°, e=30°) of powdered lunar meteorite MAC88105 at 80°C under vacuum
Type(s)
laboratory
Number of spectra
1
Instrument
SHINE Spectro-Gonio Radiometer
Sample holder
Aluminum plate with 1 mm deep circular hole (3 cm) inside P-T controled chamber closed by saphire window and cooled by Peltier elements (SERAC environmental chamber)
Standard medium
vacuum
Spectral range type(s)
Vis, NIR
Valid spectral range(s)
Min - Max (${\mu}m$) Sampling (${\mu}m$) Resolution (${\mu}m$) Position accuracy (${\mu}m$) Absorption edge
#1 0.35 - 0.67 0.02 0.0048 0.002
#2 0.67 - 1.49 0.02 0.0095 0.002
#3 1.49 - 2.99 0.02 0.019 0.002
#4 2.99 - 4.58 0.02 0.0386 0.002
Filters
Type Center/Edge (${\mu}m$) Width (${\mu}m$) Place
#1 long-pass filters of the spectro-gonio radiometer SHINE None None output of monochromator
Comments
long-pass filters to remove higher orders of the monochomator
Incidence angle(s)
0.0°
Emergence angle(s)
30.0°
Azimuth angle(s)
0.0°
Resolution illumination
0.1°
Resolution observation
4.1°
Comments
fixed angle
Illumination
Observation
Comments
partial polarization variable with wavelength (from monochromator grating)
Observation mode
single spot
Image size
23.1 x 20.0 $mm$
Spatial resolution
21.5 ${\pm}$ 1.5 $mm$
Comments
elliptic observation spot
Date begin